Care and Maintenance of Reference Electrodes

Reference electrodes are an essential component of electrochemical measurement, characterized by its stable and known reference potential. It serves as a stable reference potential for accurate measurement in wide range of electrochemical system.

It is crucial to prioritize the appropriate maintenance and handling of reference electrodes to ensure the ongoing precision of measurements. Regular and maintenance of reference electrode also ensures reproducible results and increase life span of electrode.

 

Table of Contents

The Crucial Role of Storage

When not in use, reference electrode should be stored in a suitable electrolyte solution, generally the same as the filling solution. The porous tip of the reference electrode should always remain wet to ensure proper ion conductivity, prevent contamination and reduce impedance which in turn gives precise results. Reference electrodes that are susceptible to photochemical reaction should be stored in dark conditions.

 

Regular Checks on Filling Solution Integrity

The filling solution of reference electrode should be periodically checked to ensure its working properly. The electrode should be filled appropriately and ensured no air bubble is trapped inside the electrode.

 

Post Experiment Cleaning

Regular cleaning after experiments is essential to avoid the accumulation of contaminants or residue on the reference electrode. Gently rinse the electrode with distilled water to remove any debris deposited on the surface

 

The Role of Bridge Tubes

To avoid contamination during corrosion testing, it is advisable to use a bridge tube. This helps prevent unwanted cross-contamination. Generally the bridge tube should be filled with electrolyte used in your cell.

 

Electrode testing

Master Reference Electrode

Master reference electrode is standard reference electrode, which serves as a benchmark against which other reference electrodes can be compared. It is important that you don't use the master electrode for any experiments.

Using Multimeter

Checking reference electrode with master reference electrode using multimeter

Ensure that the multimeter is set to appropriate voltage measurement range (typically millivolts mV or Volts V). Connect one end of the probe to the reference electrode being tested and the other probe to the master reference electrode. If the potential difference is around ±3mV then the reference electrode is working properly.

Using Potentiostat

Checking reference electrode with master reference electrode using potentiostat

Conduct the OCP (Open Circuit Potential) experiment using a potentiostat to measure the potential difference between the reference electrode and the master reference electrode

Connect white “reference” connector to the master reference electrode and green “working” connector to the reference electrode. The observed potential vs time graph should exhibit a relatively stable and horizontal trend (no slope) with values around ±3mV.

Conclusion:

To sum up, taking good care of reference electrodes is super important for scientists in electrochemistry. Following these guidelines helps make reference electrodes work better, giving researchers dependable and repeatable results in their experiments. Whether it's storing them right, managing filling solutions, doing proper cleaning, or doing testing, all these things keep reference electrodes in great shape, making sure they last long and provide accurate measurements in electrochemical studies.

Back to blog

1 comment

Thank you for sharing your great comments!
I wonder what solution I should use to fill in the bridge tube (for reference electrode). Would it be fine to fill the bridge tube with a filling solution of the reference electrode? Or is it possible to fill it in with electrolytes in the cell?
Your comments would be of great help. Thank you.

Jeongah Lee

Leave a comment

Please note, comments need to be approved before they are published.